SL-FL29 Oxygen Bomb Calorimeter
Ossiġnu Bomb Calorimeter / Buliding Materjali Ħruq Valur Kalorifiku Instron
introduzzjoni tal-prodott tal-ossiġnu 'Bomb Calorimeter
Huwa għal materjali tal-bini Kriterju tal-valur kalorifiku kombustjoni. Il-metodu tat-test speċifikat f'dan l-istandard jintuża sabiex jiġu ttestjati l-valur tat-tisħin assolut tal-kombustjoni tal-oġġett tat-test irrispettivament mill-morfoloġija ta 'l-artikolu
istandards
ISO1716, EN ISO 1716
Speċifikazzjoni tal-Prodott tal- ossiġnu 'Bomb Calorimeter
Bomb Kapaċità | (300 ± 50) mL |
Ossiġnu firxa ta 'pressjoni bomba | 0.5 ~ 3.5MPa, ebda tnixxija, pressjoni bomba 21Mpa |
crucible | Φ25mm, għoli (14-19) mm, ħxuna tal-ħajt ta '1mm. Stainless steel |
enerġija tħawwad | 0.5KW |
ambjent applikazzjoni | (10 ~ 35) ℃, umdità ≤80% |
kapaċità tas-sħana | dwar 10000J / K |
Repetibbiltà tal-kapaċità tas-sħana | ≤0.2% |
sejbien temperatura | (4.5-50) ℃, 0.0001 ℃ riżoluzzjoni |
sensur tat-temperatura | PT100 |
ċilindru ġewwa | 2000ml |
Ossiġnu Siġillar bomba | pressjoni ossiġinazzjoni 2.5 ~ 3.5MPa, l-ebda tnixxija |
enerġija tħawwad | 3W, 375r / min |
provvista ta 'enerġija | AC220V ± 10% 50Hz |
Prodott Karatteristiċi ta 'l ossiġnu 'Bomb Calorimeter
- Automatic control of the outer cylinder water temperature at the set point temperature, to ensure that changes in real temperature does not affect the cooling correction, the measurement results more accurate
- Automatically adjust the temperature inside and outside the cylinder, the use of a unique cooling correction system, water circulation system and software error correction system to reduce the cooling correction, the measurement results more accurate
- Meet the international standard "end of the barrel than the outer cylinder about 1K"
- computer calorific value calorimeter, to maintain the full functionality of the microcomputer system can run common software for other transactions, and start calorimeter measurement system can automatically calibrate the calorimeter energy equivalent (heat capacity), measure the calorific value
- Computerized combustion calorimeter The inner cylinder of the device adopts the electric stirring of the paddle blade, and the mixing of the outer cylinder adopts the submersible electric stirring to make the stirring more even and convenient. The instrument uses a blown cotton ignition method.
- microcomputer calorific value tester operating in WinsowsXP and above operating system, the whole process of Chinese characters prompts, human-computer interaction, that is, learning to use, follow the prompts to complete the test.
Ikteb messaġġ tiegħek hawn u jibagħtu lill magħna