EN ISO 9239-1 Radiant Flooring Panel Test Apparatus
Radiant Flooring Panel Test Apparatus Stainless Steel Hood with Smoke Measurement Ports
Product parameters (specifications)of the the Radiant Flooring Panel Test Apparatus
mana whakahaere mōrahi | 5kw |
tika wā | <1s / h |
rere Heat mita ine whānui | (0-15) Kw / m2; (Rere wera kiritaki-fatuhia) |
Heat tika mita rere | ± 0.2Kw / m2 |
Heat tika mita rere | <± 3%; |
Heat ngā iraruke pyrometer | 400-600 ℃, te tika o ± 0.5 ℃; |
hapa pāmahana tinana Black | ± 5 ℃; |
tere paowa | (2.5 ± 0.2) m / s |
kaha paowa | (39-85) m3 / min |
Maiengi rahi poari radiation whakangutungutu | 300mm × 450mm, the ability of about 900 ℃ high temperature; |
Thermocouple | diameter o thermocouple 3.2mm K-momo kowiri tira Armoured |
He wera rama | diameter ¢ diameter 10mm ¢ 6mm, rehu ngongo i runga i te poka 35 ¢ 0.7mm |
teitei mura | 60 ~ 120mm te whakarite |
pūnaha Ignition | ngaohiko nui pātīmata hiko |
He wera rahi pouaka | L1420 * W1200 * H2300 (mm) |
pūnaha whakapau kaha | rawa te mīhini e ki te mea powhiriwhiri kaha, ka oti te whakamātautau te, tīmata i te pūnaha whakapau tahi ka waiho waho ruha |
Introduction Hua o te hekenga Flooring Paewhiri Test taputapu
Te ISO 9239 hekenga Paewhiri Flooring whakamātautau ko te kokonga o te tauhohenga rauemi kuaha ki ngā whakamātautau ahi i roto i te pūnaha Euroclass. ine te whakamātautau te horapa o mura o te tauira, i raro i waiariki te rongo, mā te whakamahi i te pūtake pātīmata wera marama. Ko te puaretanga waiariki ko e o te ahi tipu i roto i te ruma o te take e roto i te kohao tatau, faaite mai ki te floorings i roto i te ruma pātata kauhanga ranei. taea hoki te whakamahi i te taputapu mō ngā whakamātautau ki ngā paerewa ASTM E648 me din 4102; Wāhanga 14.
Use Environment
Rō whare whakamahi
Altitude: ake ki 2000m
Ambient pāmahana: 10 ki te 35 ° C (60 ki te 85 ° F)
haumākū pātahi: 45% ~ 75%.
Rerekē ngaohiko tukua: 10% o te ngaohiko tauanga.
Whakatauranga Overvoltage: II
Pollution tohu: 1
Āhuahira Hua me te Tono o te Radiant Flooring Panel Test taputapu
Kiriweti 1.The ruma hua whakamātautau i te konupūmā magnesium poari mātao me te whakaratohia ki te matapihi e kitea te ātete pāmahana.
2.The stainless steel sample support assembly is mounted on a sliding platform to allow safe and easy loading of the test sample.
3.The radiant heat is applied by means of a gas-fuelled panel, inclined at 30º, and directed at a horizontally mounted floor covering system specimen.
4.The radiant panel generates a radiant energy flux distribution ranging from a nominal maximum of 10.9 kW/m2 to a minimum of 1.1 kW/m2.
5.Dummy calibration specimen with holder, calibrated heat flux meter and mounting.
6.The distance burned until flame-out is reached and converted, by calibration, into an equivalent critical radiant flux, in kW/m2.
7.Heat Flux Meter: Range : 0 ~ 50 kw/m2; Surface emissivity :ε=0.95± 0.05
8.A smoke measuring system, according to DIN 50055, is mounted on a separate frame at the exhaust stack.
9.Control Rack for convenience in use, allowing observation of the apparatus and controls during equipment set-up and calibration.
10.Automatic ignition of the radiant panel and safety cut-out.
11.Automatic moving T type ignition burner
12.Stainless steel hood with smoke measurement ports.
Taupānga: Ko te taputapu FRP Flooring hekenga Paewhiri Test ko te mea e taea te whakamahi ki te whakatau i te radiation waiariki ōrite o whariki, rauemi kuaha, me ngā rauemi mātao tuanui e e rerekë horapa i roto i te mura.
Product Details of the Radiant Flooring Panel Test taputapu
tono te hei whakamatau radiation waiariki mō te faaineine rauemi e ki te tino puia inenga radiation rere o te whenua hipoki rauemi. E pā ana te rere marama tino ki te wera raruraru ia rohe wae. Ko te rahi te wera maiki, te teitei te nui whakawera e hiahiatia mo te rauemi, me te ake eu te rauemi. Ko te iti pea ko reira ki te hopu ahi. I roto i te pūrere whakamātautau, he puna radiation wera me te T-momo pūwera e tāuta, me te atu tata ko te tauira i te pūtake radiation, te raro te radiation wera rere riro i runga i te mata, mura te tauira e te T-momo pūwera, ka whakamana i te whakamātautau te. A, no te keto te mura te, te tawhiti kino o te tauira, te hāngai uara rere radiation wera, e ko, te rere radiation wera tino.
Tutuki ana te taputapu FRP Flooring Paewhiri hekenga Test nga tikanga whakamātautau e whai ake nei:
ISO 9239-1:Floor coverings - Fire reaction test - Part 1: Determination of flaming performance using radiant heat sources;
ASTM E648:The floor covering is subjected to a critical thermal radiation test using a thermal radiation source method;
GB/T 11785:Determination of the burning properties of flooring materials - radiant heat source method.
Dimension:1900mm (W) x 750mm x (H) x 2120mm (D)
Taumaha: 235kg
whakaritenga tāutanga
Rahi Tāuta (mm)
rahi pouaka Test :: 1800 (L) * 1500 (W) * 2400 (H)
Electric rahi rūnanga mana: 600 (L) * 1000 (W) * 1800 (H)
whakaritenga tāutanga
Supply Power: 220 V±10%,50 Hz,5A
whakaritenga Air pūtake
Te tahu hau: pōwaro paerewa, he kore e iti iho i 5l.Tena / min te auau rere supply, a kahore iti ko atu 0.15MPa te pēhanga hau kokoru;
Compressed air: the air force of the intake air is not less than 0.5MPa, and the flow rate is not less than 180L/min;
Water source: dedicated coolant for cooling the heat flow meter.
Electrical: 230 volts-Ingoa 10 Amps
Hauhā Temperature: Whakahaere 10 ° C ki te 35 ° C
Dimensions: Apparatus: 1900mm (W) x 750mm x (H) x 2120mm (D)
Gas Supplies: Propane me rangi compressor
whakaritenga pūnaha ākihau
auri pape: ërä 30hm, tere hau 2.5m / s, teitei i te whenua, ko te 2.4m te.
What is the ISO 9239-1:2010?
ISO 9239-1:2010 specifies a method for testing the upwind combustion behavior and flame spread of horizontally installed floors exposed to heat flux radiation gradients in a laboratory, when ignited with a pilot flame.
This method works for all types of floors. Examples include textile carpets, cork, wood, rubber and plastic coverings and coatings. The results obtained by this method reflect the performance of the floor, including any substrates (if used). Modifications to the backing, adhesion to the substrate, other changes to the underlayment or floor may affect the test results.